7-11
Module Parameter
Low alarm limit High alarm limit
Auto alarm limits
range
Adult/
pediatric
Neonate
Adult/
pediatric
Neonate
CO
2
EtCO
2
0 to 32mmHg:
remains the
same
0 to
32mmHg:
remains the
same
0 to 32mmHg:
remains the
same
0 to
32mmHg:
remains the
same
Same as the
measurement range
32 to
35mmHg:
29mmHg
32 to
35mmHg:
29mmHg
32 to
35mmHg:
41mmHg
32 to
35mmHg:
41mmHg
35 to
45mmHg:
(etCO
2
-6)
mmHg
35 to
45mmHg:
(etCO
2
-6)
mmHg
35 to
45mmHg:
(etCO
2
+6)
mmHg
35 to
45mmHg:
(etCO
2
+6)
mmHg
45 to
48mmHg: 39
mmHg
45 to
48mmHg:
39 mmHg
45 to
48mmHg: 51
mmHg
45 to
48mmHg:
51 mmHg
>48mmHg:
remains the
same
>48mmHg:
remains the
same
>48mmHg:
remains the
same
>48mmHg:
remains the
same
FiCO
2
N/A N/A
Same as the
default alarm
limit
Same as the
default
alarm limit
Same as the
measurement range
awRR
awRR × 0.5 or
6 rpm
(whichever is
greater)
(awRR – 10)
or 30 rpm
(whichever
is greater)
awRR × 1.5 or
30 rpm
(whichever is
smaller)
(awRR+25)
or 85 rpm
(whichever
is smaller)
Adult/pediatric: 6 to
55
Neonate: 10 to 90
C.O. BT
Adult:
(BT – 1)ć
N/A
Adult:
(BT – 1)ć
N/A
Same as the
measurement range

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz