EFA-200/-300 11 Remote control
11-5
Error message: “Q<n> FAILED - wrong selected”
The data sets selected for data export are unavailable or the selection was
not made in ascending order of data set number.
Check the entries.
Error message: “Q<n> FAILED - not implemented yet”
The command is not implemented.
Error message: “Q<n> FAILED - not enough memory”
There is not enough free memory to record a timer-controlled series of
measurements.
Delete all or part of the data memory contents (see C222 or C224).
Error message: “Q<n> FAILED - mode not available“
The operating mode is not enabled.
Enabling operating modes: see page 10-9
Error message: “Q<n> FAILED - improper device type“
The setting is being transferred to an improper device type.
Connect the proper device type (Basic Unit or E-Field Unit).
Error message: “Q<n> FAILED - no standard available“
No limit curve implemented.
Load the limit curves.
Software update: see page 10-4
Error message: “Q<n> FAILED - matrix empty“
The data set cannot be saved because no measured values were saved in
the data pool.
Restart the measurement.