11 Remote control EFA-200/-300
11-10
Spectrum FFT mode
The “Run” state should be activated.
Select / read BAND STOP filter (for definition see filter frequencies)
C152 <p> C153 0 to 5 Consecutive
number within the
available list.
Select / read detection mode
C144 <p> C145 0 = RMS
1 = VECT PEAK
Write/action
commands
Read commands Range of p Meaning or
Unit of p
Select/read frequency range (see enabling of frequency range extension)
C160 <p> C161 0 = 5 Hz..2 kHz
1 = 40 Hz..32 kHz
Display spectrums
C156 <p> C157 0 = absolute
1 = normalized to
standard
Select/read detection mode
C158 <p> C159 0 = ACT RMS
1 = AVG RMS
2 = VECT PEAK
Select/read averaging parameter
C120 <p> C121 0 = 4
1 = 8
2 = 16
3 = 32
NAVG
Table 11-4 Remote control commands, Spectrum FFT mode
Write / action
commands
Read commands Range of values of p Meaning or units
of p
Table 11-3 Remote control commands, Field Strength mode