EFA-200/-300 12 Specifications
12-3
Dynamic range limit
Basic Unit
The maximum magnetic flux density that can be measured is twice the
measurement range setting. The frequency-dependent overload limit as
indicated in figure 12-1 must also be taken into account.
Example:
The dynamic range limit is 2 mT in the 1 mT measurement range when the
100 cm
2
probe is used (trace 1). The available dynamic headroom is
reduced above a frequency of 4000 Hz as shown in Fig. 12-1.
Figure 12-1 Dynamic range (for measurement range 31.6 mT)
Trace 1: 100 cm
2
probe
Trace 2: Built in probe or 3 cm probe
E-Field Unit
The maximum field strength that can be measured is twice the
measurement range setting.
Line 1
Line 2