7 Measurement data memory EFA-200/-300
7-4
Opening the AUTO menu
Date and time have been set correctly.
Setting the date and time: see page 8-9.
Setting the date and time: see page 8-9
1. Press the MEM key.
The MEM menu opens.
2. Use the UP/DOWN keys to select the “AUTO” function.
3. Press ENTER to confirm your selection.
The MEM AUTO menu is displayed.
Figure 7-2 MEM AUTO menu
Selecting the measurement sequence recording method
There are two ways to record the measured values.
By defining a time period for saving data: The precise start and stop
times for the measurement are entered.
Immediate start: The measurement is started immediately and runs until
the data memory is full (max. 24 hours).
The instrument switches off automatically as soon as the final measured
value has been saved.
Note: Start and stop times must be entered in the same format as that
selected for the clock display (12- or 24-hour clock).