Copyright© Orphée SA. All Rights Reserved.
MESSAGE ACTION
Switch on the printer or invalidate the printings.
Add some paper.
Switch on the printer or invalidate the printings.
Switch on the printer or invalidate the printings.
Switch on the printer or invalidate the printings.
Switch on the printer or invalidate the printings.
INTERN: RESULT AREA IS LOCKED.
Re-start the Mythic. Check FIFO mode. Reload last SW.
Re-start the Mythic. Check FIFO mode. Reload last SW.
Re-start the Mythic. Check FIFO mode. Reload last SW.
N/A. Analysis Mode selected is unknown. This could happen on
SVM mode.
BACKUP: NO MEMORY AVAILABLE FOR STORAGE
Delete the stored results. Check FIFO mode.
BACKUP: MEMORY IS ALMOST FULL. PLEASE
DELETE RESULTS.
Delete the stored results. Check FIFO mode.
BACKUP: BAD FOLDER DUPLICATION
Re-start the Mythic.
BACKUP: CALIBRATION HISTORY IS FULL
Delete the calibration results. Save on USB before if needed.
BACKUP: REPEATABILITY HISTORY IS FULL
Delete the repeatability results. Save on USB before if needed.
BACKUP: QC HISTORY IS FULL
Delete the Q.C. results of the ongoing lot. Save on USB before
if needed.
INTERN: NO MEMORY AVAILABLE
Communication error retry.
Re-start the Mythic. Check host connection speed.
Re-start the SVM.
Re-enter the codes or confirm it (manual connection on the
SVM).
Westgard alarm.
XB alarm.
SVM: ILLEGAL SERIAL NUMBER.
This MYTHIC cannot be connected to the SVM (wrong codes).
ID AND/OR PID MANDATORY (CHECK SETUP).
SID ALWAYS MANDATORY.
Enter an ID and/or PID and SID.
Pressure out of ranges. Check fluidic. Check vacuum syringes.
PRESSURE DEFAULT : NEEDLE RINSING DEFAULT
Check rinsing head and rinsing head fluidics. Waste syringes.
PRESSURE DEFAULT : DRAIN BATH DEFAULT
Check the tubing connection of the fluidics waste . Valves 1 & 2.
PRESSURE DEFAULT : VACUUM COUNTING
DEFAULT
Perform a piston greasing and check the tubing connection of
the fluidics. Check vacuum pressure value (section 9.3).
PRESSURE DEFAULT : NO REAGENT
Check the reagents level. Reagents tubing.
PRESSURE DEFAULT : SYRINGE VACUUM
Perform a piston greasing and check the tubing connection of
the fluidics. Check vacuum pressure value (section 9.3).
PRESSURE DEFAULT : WASTE DRAINING
Check the tubing connection of the fluidics waste. Valves 1 & 2.
BACKUP : FOLDER NOT FOUND
Check USB key and presence of correct sub-folder (see 3.4.7).
BACKUP: FILE SYSTEM FAILED.
Re-start the Mythic. Update SW to last version.
Hardware failure on memory.
Rebuild cycles with good options.
HARDWARE: F.S.C BOARD ID FAILED.
Check the hardware connection FSC Board to opt. bench.