9. SERVICE
REF : M22/UM/EN/014
Page 100/105
Copyright© Orphée SA. All Rights Reserved.
MYTHIC 22
MESSAGE ACTION
PRINTER IS BUSY
Switch on the printer or invalidate the printings.
PRINTER: NO PAPER
Add some paper.
NO PRINTER SELECTED
Switch on the printer or invalidate the printings.
PRINTER IS OFF
Switch on the printer or invalidate the printings.
NO PRINTER RESPONSE
Switch on the printer or invalidate the printings.
PRINTER ERROR
Switch on the printer or invalidate the printings.
INTERN: RESULT AREA IS LOCKED.
Re-start the Mythic. Check FIFO mode. Reload last SW.
BACKUP: SYSTEM ERROR
Re-start the Mythic. Check FIFO mode. Reload last SW.
HARDWARE: SYSTEM ERROR
Re-start the Mythic. Check FIFO mode. Reload last SW.
ANALYSIS MODE UNKNOWN
N/A. Analysis Mode selected is unknown. This could happen on
SVM mode.
BACKUP: NO MEMORY AVAILABLE FOR STORAGE
Delete the stored results. Check FIFO mode.
BACKUP: MEMORY IS ALMOST FULL. PLEASE
DELETE RESULTS.
Delete the stored results. Check FIFO mode.
BACKUP: BAD FOLDER DUPLICATION
Re-start the Mythic.
BACKUP: CALIBRATION HISTORY IS FULL
Delete the calibration results. Save on USB before if needed.
BACKUP: REPEATABILITY HISTORY IS FULL
Delete the repeatability results. Save on USB before if needed.
BACKUP: QC HISTORY IS FULL
Delete the Q.C. results of the ongoing lot. Save on USB before
if needed.
INTERN: NO MEMORY AVAILABLE
Re-start the Mythic.
INTERN: MEMORY CORRUPTED
Re-start the Mythic.
COM: CRC CONTROL ERROR
Communication error retry.
RS232: TIME OUT
Re-start the Mythic. Check host connection speed.
RS232: INTERNAL ERROR
Re-start the Mythic.
RS232: SYNCHRO ERROR
Re-start the Mythic.
RS232: ACK ERROR
Re-start the Mythic.
SVM: COM. TIME OUT
Re-start the SVM.
SVM: UNMATCH
Re-enter the codes or confirm it (manual connection on the
SVM).
SVM: WG
Westgard alarm.
SVM: XB
XB alarm.
SVM: BAD VERSION
Update the SVM software.
SVM: ILLEGAL SERIAL NUMBER.
This MYTHIC cannot be connected to the SVM (wrong codes).
ID AND/OR PID MANDATORY (CHECK SETUP).
SID ALWAYS MANDATORY.
Enter an ID and/or PID and SID.
PRESSURE DEFAULT
Pressure out of ranges. Check fluidic. Check vacuum syringes.
PRESSURE DEFAULT : NEEDLE RINSING DEFAULT
Check rinsing head and rinsing head fluidics. Waste syringes.
PRESSURE DEFAULT : DRAIN BATH DEFAULT
Check the tubing connection of the fluidics waste . Valves 1 & 2.
PRESSURE DEFAULT : VACUUM COUNTING
DEFAULT
Perform a piston greasing and check the tubing connection of
the fluidics. Check vacuum pressure value (section 9.3).
PRESSURE DEFAULT : NO REAGENT
Check the reagents level. Reagents tubing.
PRESSURE DEFAULT : SYRINGE VACUUM
Perform a piston greasing and check the tubing connection of
the fluidics. Check vacuum pressure value (section 9.3).
PRESSURE DEFAULT : WASTE DRAINING
Check the tubing connection of the fluidics waste. Valves 1 & 2.
BACKUP : FOLDER NOT FOUND
Check USB key and presence of correct sub-folder (see 3.4.7).
BACKUP: FILE SYSTEM FAILED.
Re-start the Mythic. Update SW to last version.
BACKUP: SECTOR FAILED.
Hardware failure on memory.
COM: BAD CYCLE MODULE
Rebuild cycles with good options.
HARDWARE: F.S.C BOARD ID FAILED.
Check the hardware connection FSC Board to opt. bench.