MESSAGE ACTION
HARDWARE: A.L.L BOARD ID FAILED.
Check the hardware connection ALL Board to opt. bench.
HARDWARE: HEAT REAGENT STOPPED.
Close the side door. Check your reagent sensor. Restart heater.
HARDWARE: HEAT ENCLOSURE STOPPED.
Check your enclosure sensor.
HARDWARE: HEAT REAGENT FAILED.
Call an Orphée’s representative.
HARDWARE: HEAT ENCLOSURE FAILED.
Call an Orphée’s representative.
HARDWARE: ENCLOSURE FAN FAILED.
Check if your temperature fan is running.
SYSTEM LOCKED HEATING FAILED
Call an Orphée’s representative. Restart heater.
TEMPERATURE OUT OF RANGE
Room temperature out of range (<18 or >34°C).
HEATING IN PROGRESS, PLEASE WAIT.
Wait for the system to reach enclosure temperature.
SET TEMPERATURE REACHED.
The instrument is ready to work.
CYCLE: HGB CHANNEL SATURATION. PLEASE RUN
STARTUP.
Run Startup Cycle (side door closed) If the problem still
occurs, Call an Orphée’s representative.
COM: SIZE ERROR.
Try again, if the problem occurs, call an Orphée’s
representative.
SYSTEM: DOWNLOADING NEW VERSION.
PLEASE WAIT
System is resetting after the version released.
USB: UNABLE TO OPEN FILE
The USB thumb drive as an incompatible tree structure for the
analyzer, try another or format it. See section 3.4.7.
USB: THUMB DRIVE IS FULL.
No memory available in USB for storage. You have to delete
some data to work. Enable FIFO mode.
USB: EMPTY FILE
The USB thumb drive as an incompatible tree structure for the
analyzer or no file present in the folder. See section 3.4.7.
USB: THUMB DRIVE IS NOT PRESENT.
Connect an USB thumb drive.
USB: WRITE PROTECTED FILE.
The USB thumb drive as an incompatible tree structure for the
analyzer, try another. Unlock security on USB key.
USB: TOO MANY FILES OPENED.
The USB thumb drive as an incompatible tree structure for the
analyzer or no file present in the folder. See section 3.4.7.
USB: DIRECTORY DOES NOT EXIST.
The USB thumb drive as an incompatible tree structure for the
analyzer or no file present in the folder. See section 3.4.7.
USB: UNABLE TO OPEN DIRECTORY.
The USB thumb drive as an incompatible tree structure for the
analyzer or no file present in the folder. See section 3.4.7.
USB: DIRECTORY IS NOT EMPTY.
Check USK key folders structures and contains.
USB: THUMB DRIVE I/O ERROR.
The USB thumb drive as an incompatible tree structure for the
analyzer or no file present in the folder. See section 3.4.7.
USB: UNABLE TO CREATE DIRECTORY.
The USB thumb drive as an incompatible tree structure for the
analyzer or no file present in the folder. See section 3.4.7.
SETUP: MODIFICATION NOT ALLOWED.
You have to be Logged with the good pass word.
BACKUP: LAST RESULT SAVED.
Memory full, next result will not be saved. You have to delete
results. Check FIFO mode.
BACKUP: FAIL TO SAVE RESULT ONTO THE
EXTERNAL STORAGE DEVICE. THE INTERNAL
MEMORY IS USED.
Check if the USB thumb drive is connected to the analyzer.
INVALID DATA FORMAT.
Retrieved data is invalid. The read date is not compatible with
the analyzer.
LOT ALREADY EXISTS. ACTION CANCELLED.
Select another lot.
NUMBER MAX. OF FILES REACHED.
PLEASE DELETE FILES.
Delete data on the selected directory.
VERSION RELEASE FAILED. THE CHOSEN
RELEASE IS NOT COMPATIBLE WITH THE
MYTHIC 22 SYSTEM.
Select the correct version.
PRINTER DRIVER UPDATE FAILED. THE CHOSEN
DRIVER IS NOT COMPATIBLE WITH THE MYTHIC
Select the right driver compatible with the Mythic.