MESSAGE ACTION
NETWARE: SERVER INIT. FAILED
Check your analyzer network configuration. See section 3.4.2.
NETWARE: CLIENT INIT. FAILED.
Check your analyzer network configuration. See section 3.4.2.
HOLDER: CLOSE DOOR TO ADJUST.
Message reserved to the service people with 22 CT.
HOLDER: ERROR WHILE OPENING.
Check if something prevents the door opening with 22 CT.
THE ROCKER IS BAD POSITIONNED. AUTO-
ADJUSTMENT FAILED.
Message reserved to the service people.
Check if something prevents the door closing. with 22 CT.
HOLDER: NO TUBE HOLDER DETECTED.
Put the tube holder. with 22 CT.
Check if something prevents the door opening. with 22 CT.
CYCLE : PUMP COMMAND FAILED.
Check the pump wiring. with 22 CT.
HARDWARE: BACK FAN FAILED.
Check if your back fan is running.
CYCLE: HGB CHANNEL SATURATION.
Perform startup or Technician has to adjust Hgb channel.
A USB THUMB DRIVE IS ALREADY MOUNTED ON
PORT 2. PLEASE REMOVE THE THUMB DRIVE
Remove one USB key.
A USB THUMB DRIVE IS ALREADY MOUNTED ON
PORT 1. PLEASE REMOVE THE THUMB DRIVE FORM
Remove one USB key.
THE USB THUMB DRIVE FILE SYSTEM IS NOT
COMPATIBLE WITH THE MYTHIC 22. EITHER
The USB thumb drive as an incompatible tree structure for the
analyzer or no file present in the folder. See section 3.4.7.
THE USB DEVICE PLUGGED IN PORT 1 IS NOT
COMPATIBLE WITH THE MYTHIC 22. PLEASE
The USB thumb drive as an incompatible tree structure for the
analyzer or no file present in the folder. See section
3.4.7.
THE USB DEVICE PLUGGED IN PORT 2 IS NOT
COMPATIBLE WITH THE MYTHIC 22. PLEASE
The USB thumb drive as an incompatible tree structure for the
analyzer or no file present in the folder. See section
3.4.7.
A USB PRINTER IS ALREADY PLUGGED IN PORT
2.PLEASE REMOVE THE PRINTER FROM THE PORT
Remove one USB printer.
A USB PRINTER IS ALREADY PLUGGED IN PORT
1.PLEASE REMOVE THE PRINTER FROM THE PORT
Remove one USB printer.
TUBE HOLDER SWITCH ERROR, CLEAN THEM AND
GO TO CHECK SENSORS MENU.
Call an Orphée’s representative.
SYSTEM: PREAMPLIFIER BOARD IS NOT
COMPATIBLE WITH YOUR SYSTEM.
Call an Orphée’s representative.
HOLDER: TUBE HOLDER NOT AUTHORIZED
Remove Tube Holder. with 22 CT.
Replace the container and perform a prime cycle.
Replace the container and perform a prime cycle.
Replace the container and perform a prime cycle.
Input correct Orphee reagent.