COMMAND REFERENCE GUIDE
Original Instructions
Example of reading a temperature sensor
The following command reads the sensor voltage:
READ:DEV:MB0 :SIG:VOLT?
The sensor returns the following message:
STAT:DEV:MB0 :SIG:VOLT:12.345:mV
10.3.4.3 Addressing a heater controller
To address a heater controller, use the following structure:
DEV:<UID>:HTR
Where <UID> is a unique identifier that is allocated to each board, based on its SPI
location.
The configuration settings for a heater controller are:
The signals for a heater controller are:
10.3.4.4 Addressing a level meter sensor
To address a level meter sensor, use the following structure:
DEV:<UID>:LVL
Where <UID> is a unique identifier that is allocated to each board, based on its SPI
location.
HTR Heater controller designator N/A
MAN Manufacturing designator N/A
HVER Hardware version Read only version number
FVER Firmware version Read only version number
SERL Serial number Read only alphanumeric
NICK Nickname Read/set alphanumeric
VLIM Voltage limit Read/set Float value
[0.0 to 40.00]
RES Nominal resistance Read/set Float value
[20.00 to 100.00]
PMAX Maximum power supported Read only Float value
VOLT Most recent voltage reading/voltage
to set
Read/set Float value
[0.0 to VLIM]
CURR Most recent current reading Read only Float value
POWR Most recent power reading/ power to
set
Read/set Float value
[0.0 to Max Power]