EasyManua.ls Logo

Oxford Instruments Mercury iTC - 3.2.12.11 Factory Settings Tab

Oxford Instruments Mercury iTC
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Handbook
November 2011 / 59-UMC0071 / Issue 01
Mercury iTC
Page 53
USER INTERFACE
Original Instructions
3.2.12.11 Factory Settings tab
The page contains the following fields and controls:
Revert to factory defaults
Tap Save to save the present configuration as a restore point. This allows the present
configuration to be restored at some future date. Each restore point is identified by an
integer.
Drop-down menu
Select a saved restore point to load. Loading a restore point changes the unit’s
configuration to the saved configuration. (NOTE: Restore point 0 is the factory
configuration).
The page contains the following buttons:
Alarm
Tap once to access the Alarm Logs page (see Section 3.2.13).

Table of Contents