EasyManua.ls Logo

Oxford Instruments Mercury iTC - Configuring External Interfaces

Oxford Instruments Mercury iTC
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Handbook
November 2011 / 59-UMC0071 / Issue 01
Mercury iTC
Page 90
BASIC OPERATING INSTRUCTIONS
Original Instructions
7 To remove the X axis scale, tap once on the bottom of the X axis (see Figure 6-2).
Figure 6-2 Signal Monitor Plot page with Y axis scale only
6.4.2 To remove a trace from the plot
1 Tap Plot on the Home page. The Plot page is displayed.
2 Double-tap anywhere on the page. The Plot Configuration page is displayed.
3 Tap Device and select the required device.
4 Tap Signal and select the trace that is to be removed from the plot.
5 Tap Delete. The selected trace is removed from the plot.
6.4.3 To exit the plot
1 Double-tap anywhere on the page. The Plot Configuration page is displayed.
2 Tap Home to return to the Home page.
6.5 Configuring external interfaces
Only one external interface can be used at a time. refer to the relevant section for
information on configuring an external interface.
6.5.1 Configuring the ethernet interface (if required)
1 Obtain the required ethernet settings from the network administrator.
2 Tap Settings on the Home page.
3 Scroll to the Ethernet tab.
4 Tap DHCP and select either On or Off.
5 If DHCP is set On, the unit obtains its ipaddress from the network. No further
configuration is required.
6 If DHCP is set Off, enter an IP Address for the unit.
7 If DHCP is set Off, enter a Subnet Mask for the unit.
8 If DHCP is set Off, enter a Gateway value for the unit.

Table of Contents