96
Cedex HiRes Analyzer Operator’s Guide
D
D
6.2.2. Import
Once measurement data has been exported from one system to external storage, the data can then be imported
to another system. To import measurement data into a Cedex Software 2.2 database, follow the instructions
below:
1
Select the Data Exchange option in the Functions menu on the Cedex Control Center.
2
Click on Import to open the Import From External DB window.
3
Enter the path of the fi le to be imported in the Location area. Alternatively, the button on the right of the
entry fi eld opens a window in which the path of the fi le can be directly selected. Click on Open. The Import
From External DB window will reappear.
4
Click on the Import button to start the import process. A message window will appear when the import is
successfully completed. Click on OK.
5
Close the Import From External DB window by clicking on .
■
It is not possible to import more than 200 measurements at one time.
A default entry for the path under which the import fi le should be stored can be entered in the System Options
under the Setup menu. Open the Data Exchange tab and look for the ImportPath parameter. In the Va l u e
column, enter the default path. (see list of User Preferences and System Options in the Appendix).
To view imported measurements that come from a different system and, consequently, have a different System
ID, select the appropriate System ID for those measurements in the SystemName dropdown menu in the
Measurement List window.
6.3 Viewing and Evaluating the Audit Trail File
The Cedex Software 2.2 automatically creates an Audit Trail (log) fi le for each sample and records all relevant
hardware information relating to the sample. The Audit Trail fi le data can never be overwritten. This ensures
that all recorded, analyzed and modifi ed measurement data is documented (see “The System Audit Trail File”).
General Functions of the System