Cedex HiRes Analyzer Operator’s Guide
C
C
5.2 Logging out
To log out of the Cedex Control Center, select Logout under the System menu or click on the Logout button in
the Cedex Control Center. If an LM Shutdown has not yet been carried out, the message, “Do you really want to
log out without performing an LM Shutdown?”, will appear. Either click on Ye s to log out immediately or click
on No, perform an LM Shutdown, then log out.
The software will be ended and the login window will reappear. Logout is possible during a measurement
without infl uencing it. A message indicates that measurements are in progress.
5.3 Shutting down the Control Center
Follow the instructions below:
1
Perform an LM Shutdown to shut down the Liquid Management system
(see “
Carrying out an LM Shutdown”).
2
Shut down the software by selecting Exit under the System menu or by clicking on to close the window.
The security query, “Do you really want to exit the Cedex Control Center?“, will appear.
3
Click on Ye s to shut down the program.
Restarting the analysis software is necessary for the next measurement session (see Starting the Cedex HiRes
Software 2.2 ”).
If at all possible, the analyzer should never be turned off. If it is necessary to turn it off, the
analyzer needs at least
2 hours to warm up before it can be used for measurements.
Switching off the Cedex HiRes Analyzer
Shutting down the Control Center