60
Cedex HiRes Analyzer Operator’s Guide
C
C
Figure 45: Adding sample information to the Multi Run window
4.3.3. Activating Sample Rows and Starting the Measurement
After all sample information has been entered in the Multi Run preparation area, the samples can be activated
and, thus, added to the process queue. To activate samples and start the measurement, do the following:
1
Activate the status of each sample row by double clicking in the Status eld of the row to be activated. The
row appears pale blue, and further modifi cations of sample information is impossible. Return to the Edit
status by double clicking on the Status area of the sample row to allow for further changes, if required.
2
Alternatively, all samples can be activated or deactivated simultaneously by double clicking on the word
Status in the column heading at the top of the Multi Run preparation area. Confi rm the safety query with
Ye s to accept the modifi cation.
3
After some or all rows have been activated, click on the Start button on the right-hand side of the Multi Run
window to start the analysis of the activated measurements.
When measuring begins, the row of the sample being measured will be highlighted in yellow, and the progress
of the measurement will be tracked in the status column. When a measurement is fi nished, the row will be
highlighted in green and status column will indicate Done. In addition, the estimated time required for the
measurement is shown in the Estimated time remaining eld.
Images taken of the sample are displayed in the upper right-hand corner of the Multi Run window during the
measurement. In addition, results for Total Cell Concentration, Viable Cell Concentration and Viability are
displayed in the upper left-hand corner of the Multi Run window during and after a measurement has fi nished.