62
Cedex HiRes Analyzer Operator’s Guide
C
C
Figure 46: Multi Run window during measurement
Only samples that have the status of active or scheduled will be measured. Samples that are in edit mode
have not yet been added to the process queue and will not be measured until they have been activated.
If an automated cleaning has been scheduled via the Scheduler function, the cleaning routine has a lower
priority than scheduled measurements and will be added to the process queue after all activated (status: ac-
tive or scheduled) measurements have been carried out (see “ Scheduling Automatic Cleaning Routines” for
more information about automating cleaning routines). Note that measurement information placed in the
Multi Run preparation area that have the status of edit are not added to the queue! Therefore, if a Cedex
Sample Cup containing a sample is added to the Multi Sample Tray in the DefaultCup position, and the row
for that sample is left in edit mode, the cup will be used for a cleaning routine should one be scheduled to be
carried out at that time. It is therefore critical to ensure that a Cedex Sample Cup containing a sample to be
measured is never placed in the DefaultCup position on Multi Sample tray and left in edit mode during a
Multi Run. Samples placed in the DefaultCup position for measurement should be immediately set to active
or scheduled in order to be added to the process queue and, thus, have precedence over all scheduled cleaning
routines.
4.3.4. Adding Measurements to the List while Measurements are Running
Sample rows in which data has not yet been entered can be fi lled out at any time while the Multi Run is running.
In addition, sample rows containing fi nished measurements (status: done) can be cleared and prepared for a
new measurement by double-clicking in the Status eld of the row with the fi nished measurement. Sample data
for a measurement cannot be modifi ed after that measurement has begun.