Administrator Functions
Adjustment of the Image Analysis using the Live Operator
7.5.7. Working with the Live Operator
The Live Operator window can be opened via the Cell Type List window, the Cell Type window and the
Measurement window. If the Live Operator button in a Measurement window containing specifi c results
is clicked on, the Live Operator will open with the images of that measurement already loaded. If the Live
Operator is opened via the Cell Type or Cell Type List windows, the fi rst measurement in the Measurement
List will be loaded.
In the Live Operator window, some of the 11 parameters can be displayed on the screen at the same time.
Parameters at fi rst not visible can be scrolled into view using the scroll bar under the Operator Parameter
display. The table below displays the positions and settings of the 11 Operator parameters. See also “
Description
of the Individual Operator Parameters” for a detailed description of each parameter.
Identifi er Minimum Maximum Min. Size Std. Size Max. Size
Classifi cationBias -20 20 -1 0 -8
CMinSize [µm] 0 100 3 7 10
LCAppearance -10 10 4 0 0
DCAppearance -10 10 0 0 -1
CMaxSize [µm] 0 100 70 70 100
AggrMinSize [µm] 8 50 12 12 15
BorderWidth 0 1000 10 10 10
AggrMaxSize [µm] 10 200 100 100 100
LCAggrApp -10 10 5 0 0
DCAggrApp -10 10 6 0 -3
MinPollutionArea [µm] 50 1000 100 100 100
Table 1: A list of the Operator parameters