69
Software
D
D
In addition, the Result Data area for the individual images has an Excluded column next to the image number.
Images that could lead to incorrect measurement results during analysis, e.g. due to impurities or large bubbles,
are automatically recognized as invalid and excluded from evaluation by the analysis software. If this occurs, the
image will be marked with a check in the checkbox of the Excluded column. Users can also manually exclude
individual images from the calculations or include images that were automatically excluded by the software by
clicking on the check-box in this column.
1.2 Sample Parameters and Processing Parameters areas
The Sample Parameters area of the Measurement window shows the parameters entered by the user before
running the measurement. If any measurements have been reprocessed, the original or reprocessed data can
be selected in the drop-down menu of the Data Set section. If data has been reprocessed at some point, the
Effective checkbox will indicate which data set has been determined to be the effective version (see “
The
Reprocess Button”). Comments for the measurement can be added in this area at any time.
The Processing Parameters area lists the process parameters of the measurement, including Cell Type, used
to analyze the images (see “
Adjustment of the Image Analysis using the Live Operator”), the precision selected
and the date and time the process was run. None of the values in this area can be changed. Click on the Show
parameters button below the Cell Type drop-down menu to open a window that shows the actual Live Operator
Parameters values used for that Cell Type (see Figure 50, below).
Figure 50: Window displaying the Live Operator parameter settings for the Cell Type used for a measurement
Using the Measurement Results Window
Sample Parameters and Processing Parameters areas