109
Software
D
D
Administrator Functions
The System Options
7.4.5. Error Level Displayed to Users
Errors in the software and hardware will be displayed to the user in the form of an error message. The
level of error that will result in the display of an error message depends on the setting of the parameter
ErrorLevelDisplayToUser in the General tab of the System Options window. The following levels are available:
Level 1 (Fatal): fatal error, system/hardware crash
Level 2 (Critical): Error as a result of invalid measurement
Level 3 (Error): Error, but the result does not cause an invalid measurement
Level 4 (Warning): Information for user.
If, for example, the ErrorLevelDisplayToUser is set at Level 1, only error messages regarding system or hardware
crashes will be displayed to users. The factory-installed default for the ErrorLevelDisplayToUser is set at Level
3 (Error).
7.4.6. Liquid Container Confi guration
In the Liquid Container Confi guration area, the minimum and maximum liquid levels for reagent and waste
containers can be set. To open the Liquid Container Confi guration window proceed as follows:
1
Click on the Master Data menu on the Cedex Control Center.
2
Click on Liquid Container Confi guration and the window will appear.
Maximum and minimum volumes can be defi ned using this window. When standard sized containers are used,
it is not advisable to change default settings for the Min volume and Max volume.
It is also possible to defi ne a Warn Level and a Lock Level. The standard default for the Warn Level is 6. This
means that a warning message will appear if the amount of reagents left is 6 ml or less. The standard setting for
the Lock Level is 3, which means that the system will be locked (i.e., no further measurements will be possible)
if there are only 3 ml left. The settings for the Warn Level and Lock Level can be changed. Saving changes will
immediately update the estimation of the number of remaining measurements.
7.4.7. Disabling the Reagent Control
The Reagent Control function keeps track of the amount of reagents used and notifi es the user when the levels
are running low. The function can be disabled in the General tab of the System Options window. The Reagent
Control function is enabled when the value for the DisableReagentControl entry is set to 0. To disable the
function, set the value to 1”.
7.4.8. Default Settings
In general, research or production staff users are not assigned access rights to the System Options window in
which default settings that apply to the whole system can be confi gured. However, some parameters needed to
customize the individual workplace (e.g. defi nition of the default settings for certain measurement parameters)
can be found under User Preferences (see User Preferences”).