33
System Description
B
B
4.2 Starting up the Software
Once the installation of the software is complete, the software can be started and fi nal adjustments made. Do
the following:
1
Check for the presence of the Innovatis MicroScanner in the Device Manager as shown below:
Figure 18
2
Start Cedex Server rst (if the Cedex Server has not already automatically started), then the Cedex Client.
3
Windows Security Alert may possibly ask whether the programs CedexServer and GUI should be blocked.
If this occurs, press the Unblock button for each.
4
Log in as SuperUser (password: spass).
5
Confi rm the empty reagent container warnings.
6
If there is a message that hardware components could not be found at the confi gured serial ports, terminate
the Cedex Client and the Cedex Server after the message Hardware startup completed appears, and
restart the software as described in step 1, above.
7
Verify that the system is not running in demo mode (i.e., make sure there is no Demo label visible in the
Cedex Control Center).
8
Open the Diagnosis window via the Setup menu of the Cedex Control Center and verify the confi guration
of the system. For a Cedex HiRes Analyzer with Multi Sampler there must be a module ASXPort (or AS20Port
if the Multi Sampler was confi gured for AS20 mode).
9
Each Cedex HiRes Analyzer has a few unique specifi cations related to the particular fl ow chamber that
has been put in. These numbers must be verifi ed and/or adjusted in the System Options window before
measurements can be run. Verify and/or make the following changes in the System Options window
under either the General or the Hardware tabs:
a. Verify MultiSamplerPresent (0 = Single Port, 1 = Auto/Multi Sampler) in the Hardware tab.
b. Enter the appropriate value for the ChamberHeight in the Hardware tab.
c. Enter the appropriate value for the FlowFactor in the General tab.
10
Replace or refi ll the Reagent Kit containers. See “Preparing the Analyzer for Measurements”.
11
Carry out a Prime.
12
Carry out an automatic adjustment via the Automatic Adjustment window in the Functions menu.
The Cedex HiRes Analyzer is now ready to use.
Installing Cedex HiRes Software 2.2 or Updating to Cedex HiRes Software 2.2
Starting up the Software