107
Software
D
D
Administrator Functions
The System Options
7.4.2. Modifying the Standard Setting for the Multi Sampler DefaultCup
The default setting for the Multi Sampler DefaultCup (Cedex Sample Cup used for Prime, LM ShutDown,
focus calibration and all cleaning routines) is position 1 on the Multi Sample tray. The default position can be
changed to any other position. To change the position for the DefaultCup location, do the following:
1
Open the System Options window and click on the General tab.
2
Look for the parameter DefaultCupPosition. The number entered in the Value column will determine the
position for the DefaultCup on the multi sampler tray.
3
Click on Save to confi rm the change.
■
The position chosen for the Multi Sampler DefaultCup does not infl uence measurements. All 20 positions on
the sampler plate can be used for measurements. Should a sample be present in the DefaultCup position during
Prime, LM ShutDown, calibration or any cleaning routines, it can no longer be used for measurements.
7.4.3. Turning off Automatic Incrementation of Sample Cup Position
In the case of single measurements run via the Measurement window on a HiRes system with a Multi Sampler
attached, the software will automatically incrementally increase the sample cup position number with each new
measurement. To switch off this feature, do the following:
1
Open the Systems Options window and click on the Measurement tab.
2
Look for the parameter SuggestNextCupPosition.
3
Select 0 from the drop-down menu to turn off the feature.
4
Click on Save to confi rm the change.
■
When 0 has been selected, the suggested cup position will always be the same.