97
Software
D
D
6.4 User Preferences
The User Preferences window is located under the Master Data drop-down fi le menu in the Cedex Control
Center (see Figure 71, below). A logged-in user with specifi c access rights to the User Preferences option can set
preferences for a number of elements including the result parameters to be displayed in the Multi Run result list
and Measurement List window and default measurement parameters for carrying out measurements.
Figure 71: User Preferences window with the tab Measurement List Default Filter selected
6.4.1. Adjusting Result Display in the Measurement List and Multi Run windows
The result data to be displayed in the Measurement List window, as well as the result list area of the Multi Run
window, can be adjusted in the User Preferences window. The numbers in the entry ResultsToShowInList
located under the Measurement List Default Filter tab indicate which types of results will be displayed in the
Measurement List. The numbers in the entry DisplayedResultTypes under the General tab indicate which
types of results will be displayed in the result list of the Multi Run window as follows:
1: Avg Diameter
5: Avg Compactness
6: Viable Cell Count
7: Dead Cell Count
8: Total Object Count
9: % Viability
10: Viable Cell Conc.
11: Dead Cell Conc.
12: Total Cell Conc.
13: Std. Deviation
14: Total Cell Count
15: % Aggregate Rate
The default entry for these fi elds is 12,10,9, which corresponds to Total Cell Conc., Viable Cell Conc. and %
Viability, respectively. The order in which the numbers are listed is the order in which the results will appear in
the lists. To add more results to be displayed, enter the number that corresponds to the desired result parameter
in the fi elds. Numbers must be separated by a comma with no space in between.
General Functions of the System
User Preferences