45
Operation
C
C
2 Preparing the Analyzer for Measurements
2.1 Switching on
Each element is switched on separately as follows:
1
Switch on the Cedex HiRes Analyzer.
2
Switch on the monitor and the PC.
■
Cedex HiRes Analyzer hardware components must be switched on before starting the analysis software.
Otherwise, the analysis software may not be correctly initialized.
2.2 Logging in at the Computer Operating System Level
A prompt for a username and password will appear on the computer when everything has been turned on
correctly. The user should log in as Win-Admin, password Win-Admin unless a different Windows operating
system account has been set up.
2.3 Starting the Cedex HiRes Software 2.2
To launch the Software, do the following:
1
Check that the Cedex Server software has already automatically started (it may appear as a minimized
window in the task bar at the bottom of the computer screen). If not, start the server by double-clicking on
the Cedex Server icon on the desktop.
2
Start the Cedex HiRes Software 2.2 by double clicking the Cedex Client icon on the desktop (the Cedex
Server must be started before the Cedex Client can be started). The Log In window of the Cedex HiRes
Analyzer program will open.
3
Enter the username and password (if a password is required) in the respective fi elds.
Factory installed default user is Superuser (password spass).
4
If you are logging in for the fi rst time under a new username, you might be forced to change your password.
5
Click on the OK button to log in or press <Return>. The Cedex Control Center appears after a successful
login.
■
The user may be prompted to create a new password the fi rst time the Cedex HiRes Software 2.2 is started. This
serves the purpose of replacing the temporary password given to the user by the Administrator when the user
account was created. Follow the program instructions or see “Changing own Password“.
The Administrator, Superuser or regular user has access to the assigned functions once he/she has logged into
the Cedex Control Center. The regular user only has access rights to routine operations depending on the
confi gurations set by the Administrator. Access rights may vary according to options chosen for each user
account.
Preparing the Analyzer for Measurements