Loading...
127
Maintenance and Care
E
E
The Current area displays the scanned image.
Since the fl ow chamber is represented as a whole, it is necessary to enlarge single regions (especially the inlet and
outlet) with the help of the zoom function in order to be able to effectively check the chamber.
Certain areas of the Scan can be enlarged by following the instructions below:
1
Place the mouse pointer on the top left-hand corner of the area to be enlarged.
2
Hold the left mouse button pressed down and move it to the lower right-hand corner of the area to be enlarged.
3
Release the mouse button and the selected area will fi ll the whole window.
Return to the original size of the scan by following the instructions below:
1
Place the mouse pointer anywhere on the diagram.
2
Hold the left mouse button pressed down and form a box from the bottom right corner to the upper left corner.
3
Release the mouse button. The scan will then return to in its original size.
The brightness and contrast of the image can also be changed for a better analysis.
If a previous cleaning routine was unsuccessful, i.e. if any (dirt) particles are visible in the Scan image, a new
cleaning of the fl ow chamber must be done. Carry out another Scan to check this cleaning. The image created
after this scan can be viewed in the Current area. The image taken previous to this scan can be seen in the
Previous area.
Figure 79: Scanning via the Cleaning window
Daily and Manual Maintenance and Cleaning Routines
Scanning the Chamber to Verify the Effectiveness of Cleaning Routines