3
Maintenance and Care
E
E
Syringe Maintenance
Section E: Maintenance and Care
Chapter 5: Syringe Maintenance
Follow the syringe maintenance procedure described on pages 130 – 131 to remove, reinstall or replace the
syringe of the Cedex HiRes System. After removal, reinstallation or replacement of the syringe, a calibration
check of the Cedex HiRes System must also be performed. Use Cedex Density Reference Standard Beads to
check the calibration and confirm that the FlowFactor is correct. If necessary, adjust the FlowFactor.
To perform the Cedex HiRes System calibration check and FlowFactor adjustment, follow the instructions
described in the online Instructions for Use for the Density Reference Standard Beads, available for download
at:
https://www.roche-applied-science.com/PROD_INF/index.jsp?&id=iforu
Ordering Information
Product Pack Size Cat. No.
Density Reference Standard Beads 1× 10 ml Batch A 06 422 659 001
Density Reference Standard Beads 1× 10 ml Batch B 06 422 667 001