Remote Commands
R&S
®
FSW
1009User Manual 1173.9411.02 ─ 19
Parameters:
<Format> GDI
Graphics Device Interface.
Default format for the output to a printer configured under Win-
dows. Must be selected for the output to the printer interface.
Can be used for the output to a file. The printer driver configured
under Windows is used in this case and a printer-specific file for-
mat is thus generated.
BMP, JPG, PNG
Data format for output to files only.
Usage: SCPI confirmed
Manual operation: See "Output Medium" on page 562
HCOPy[:IMMediate<device>]
This command initiates a print job.
If you are printing to a file, the file name depends on MMEMory:NAME.
Suffix:
<device>
.
1 | 2
Printing device.
Usage: Event
SCPI confirmed
Manual operation: See "Printing or Storing a Screenshot (Print Screen)"
on page 561
HCOPy[:IMMediate<device>]:NEXT
This command initiates a print job.
If you are printing to a file, the file name depends on MMEMory:NAME. This command
adds a consecutive number to the file name.
Suffix:
<device>
.
1 | 2
Printing device.
Usage: Event
Manual operation: See "Printing or Storing a Screenshot (Print Screen)"
on page 561
HCOPy:ITEM:ALL
This command includes all screen elements in the printout.
The screen elements include comments, title, time and date.
Usage: