Remote Commands
R&S
®
FSW
1037User Manual 1173.9411.02 ─ 19
SYSTem:IDENtify[:STRing] <String>
This command defines the response to *IDN?.
Parameters:
<String> String containing the description of the instrument.
Manual operation: See "Identification String" on page 651
SYSTem:LXI:INFo?
This command queries the LXI settings.
Return values:
<LXIInfo> <current version> | <LXI class> | <Computername> |
<MAC adress> | <IP adress> | <Auto MDIX>
String containing the current LXI parameters.
• <version>
• <LXIClass>
• <ComputerName>
• <MACAddress>
• <IPAddress>
• <AutoMDIX>
Usage: Query only
Manual operation: See "Current LXI Configuration" on page 656
SYSTem:LXI:LANReset
This command resets the LAN configuration as required by the LXI standard. The com-
mand also resets the LXI password and instrument description.
Usage:
Event
Manual operation: See "LAN Reset" on page 657
SYSTem:LXI:MDEScription <Description>
This command defines the LXI instrument description.
Parameters:
<Description> String containing the instrument description.
Manual operation: See "LXI Manufacturer Description" on page 657
SYSTem:LXI:PASSword <Password>
This command defines the LXI password.
Parameters:
<Password> String containing the password.
Configuring the R&S FSW