EasyManua.ls Logo

Rohde & Schwarz FSW26 - Page 126

Rohde & Schwarz FSW26
1150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurements
R&S
®
FSW
126User Manual 1173.9411.02 ─ 19
Fig. 5-8: Measurement of two equally-leveled RF sinusoidal signals with a resolution bandwidth
(1
kHz) which is significantly smaller than their frequency spacing
5.2.4.2 Measuring the Modulation Depth of an AM-Modulated Carrier in the Frequency
Domain
In the frequency range display, the AM side bands can be resolved with a narrow
bandwidth and measured separately. The modulation depth of a carrier modulated with
a sinusoidal signal can then be measured. Since the dynamic range of a signal ana-
lyzer is very large, extremely small modulation depths can also be measured precisely.
For this purpose, the R&S FSW provides measurement routines that output the modu-
lation depth numerically in percent directly.
Signal
Generator
R&S FSW
Fig. 5-9: Test setup
Table 5-4: Signal generator settings (e.g. R&S SMU)
Frequency 128 MHz
Level -30 dBm
Modulation 50 % AM, 10 kHz AF
Basic Measurements