Common Measurement Settings
R&S
®
FSW
326User Manual 1173.9411.02 ─ 19
For details on the zeroing process refer to "How to Zero the Power Sensor"
on page 330.
Remote command:
CALibration:PMETer<p>:ZERO:AUTO ONCE on page 895
Frequency Manual
Defines the frequency of the signal to be measured. The power sensor has a memory
with frequency-dependent correction factors. This allows extreme accuracy for signals
of a known frequency.
Remote command:
[SENSe:]PMETer<p>:FREQuency on page 898
Frequency Coupling
Selects the coupling option. The frequency can be coupled automatically to the center
frequency of the instrument or to the frequency of marker 1.
Remote command:
[SENSe:]PMETer<p>:FREQuency:LINK on page 898
Unit/Scale
Selects the unit with which the measured power is to be displayed. Available units are
dBm, dB, W and %.
If dB or % is selected, the display is relative to the reference value that is defined with
either the "Meas -> Ref" setting or the "Reference Value" setting.
Remote command:
UNIT<n>:PMETer<p>:POWer on page 901
UNIT<n>:PMETer<p>:POWer:RATio on page 901
Meas Time/Average
Selects the measurement time or switches to manual averaging mode. In general,
results are more precise with longer measurement times. The following settings are
recommended for different signal types to obtain stable and precise results:
"Short"
Stationary signals with high power (> -40dBm), because they require
only a short measurement time and short measurement time provides
the highest repetition rates.
"Normal"
Signals with lower power or modulated signals
"Long"
Signals at the lower end of the measurement range (<-50 dBm) or
Signals with lower power to minimize the influence of noise
"Manual"
Manual averaging mode. The average count is set with the Average
Count (Number of Readings) setting.
Remote command:
[SENSe:]PMETer<p>:MTIMe on page 899
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe] on page 899