Remote Commands
R&S
®
FSW
743User Manual 1173.9411.02 ─ 19
<PClass> 1..4
Power class for which the limit is defined.
Parameters:
<Level> Relative limit at the stop frequency of a SEM range.
Range: -400 to 400
*RST: -50
Default unit: dBc
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP -15
For a detailed example see chapter 11.5.6.11, "Example: SEM
Measurement", on page 762.
Manual operation: See "Rel Limit Start/Stop <n>" on page 213
[SENSe:]ESPectrum<sb>:RANGe<range>:LIMit<PClass>:RELative:STOP:
ABSolute <Limit>
This command defines an absolute limit for the MAX function of the relative limit for a
SEM range.
For more information see "Relative limit line functions" on page 202.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the SEM range.
<PClass> 1..4
Power class for which the limit is defined.
Parameters:
<Level> Absolute limit at the stop frequency of a SEM range to be used
in addition to the relative limit if the MAX function is enabled (see
[SENSe:]ESPectrum<sb>:RANGe<range>:
LIMit<PClass>:RELative:STOP:FUNCtion
on page 744).
Range: -400 to 400
*RST: -13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
ABSolute -15
For a detailed example see chapter 11.5.6.11, "Example: SEM
Measurement", on page 762.
Manual operation: See "Rel Limit Start/Stop <n>" on page 213