Remote Commands
R&S
®
FSW
744User Manual 1173.9411.02 ─ 19
[SENSe:]ESPectrum<sb>:RANGe<range>:LIMit<PClass>:RELative:STOP:
FUNCtion <Function>
This command enables the use of a function when defining the relative limit for a SEM
range.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the SEM range.
<PClass> 1..4
Power class for which the limit is defined.
Parameters:
<Function> Defines the function to be used to determine the relative limit
line stop value
MAX
The maximum of the relative and the absolute level is used as
the limit stop value. Use the [SENSe:]ESPectrum<sb>:
RANGe<range>:LIMit<PClass>:RELative:STOP and
[SENSe:]ESPectrum<sb>:RANGe<range>:
LIMit<PClass>:RELative:STOP:ABSolute commands to
define these values.
OFF
No function is used, the relative limit line is defined by a fixed
relative stop value. Use the [SENSe:]ESPectrum<sb>:
RANGe<range>:LIMit<PClass>:RELative:STOP com-
mand to define this value.
*RST:
OFF
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
FUNCtion MAX
For a detailed example see chapter 11.5.6.11, "Example: SEM
Measurement", on page 762.
Manual operation: See "Rel Limit Start/Stop <n>" on page 213
[SENSe:]ESPectrum<sb>:RANGe:LIMit<PClass>:STATe <State>
This command selects the limit check mode for all SEM ranges.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<PClass> 1..4
Power class for which the limit is defined.
Configuring and Performing Measurements