Remote Commands
R&S
®
FSW
872User Manual 1173.9411.02 ─ 19
Band Frequency start [GHz] Frequency stop [GHz]
D 110.0 170.0
G 140.0 220.0
J 220.0 325.0
Y 325.0 500.0
USER 32.18
(default)
68.22
(default)
*) The band formerly referred to as "A" is now named "KA".
[SENSe:]MIXer:HARMonic:HIGH:STATe <State>
This command specifies whether a second (high) harmonic is to be used to cover the
band's frequency range.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
MIX:HARM:HIGH:STAT ON
Manual operation: See "Range 1/2" on page 369
[SENSe:]MIXer:HARMonic:HIGH[:VALue] <HarmOrder>
This command specifies the harmonic order to be used for the high (second) range.
Parameters:
<HarmOrder numeric value
Range: 2 to 61 (USER band); for other bands: see band
definition
Example:
MIX:HARM:HIGH 2
Manual operation: See "Harmonic Order" on page 369
[SENSe:]MIXer:HARMonic:TYPE <OddEven>
This command specifies whether the harmonic order to be used should be odd, even,
or both.
Which harmonics are supported depends on the mixer type.
Parameters:
<OddEven> ODD | EVEN | EODD
*RST: EVEN
Example:
MIX:HARM:TYPE ODD
Manual operation: See "Harmonic Type" on page 369
Setting Basic Measurement Parameters