Remote Commands
R&S
®
FSW
873User Manual 1173.9411.02 ─ 19
[SENSe:]MIXer:HARMonic[:LOW] <HarmOrder>
This command specifies the harmonic order to be used for the low (first) range.
Parameters:
<HarmOrder> numeric value
Range: 2 to 61 (USER band); for other bands: see band
definition
*RST: 2 (for band F)
Example:
MIX:HARM 3
Manual operation: See "Harmonic Order" on page 369
[SENSe:]MIXer:LOSS:HIGH <Average>
This command defines the average conversion loss to be used for the entire high (sec-
ond) range.
Parameters:
<Average> numeric value
Range: 0 to 100
*RST: 24.0 dB
Default unit: dB
Example:
MIX:LOSS:HIGH 20dB
Manual operation: See "Conversion loss" on page 369
[SENSe:]MIXer:LOSS:TABLe:HIGH <FileName>
This command defines the file name of the conversion loss table to be used for the
high (second) range.
Parameters:
<FileName> string ('<file name>')
Example:
MIX:LOSS:TABL:HIGH 'MyCVLTable'
Manual operation: See "Conversion loss" on page 369
[SENSe:]MIXer:LOSS:TABLe[:LOW] <FileName>
This command defines the file name of the conversion loss table to be used for the low
(first) range.
Parameters:
<FileName> string ('<file name>')
Example:
MIX:LOSS:TABL 'mix_1_4'
Specifies the conversion loss table mix_1_4.
Manual operation: See "Conversion loss" on page 369
Setting Basic Measurement Parameters