Remote Commands
R&S
®
FSW
874User Manual 1173.9411.02 ─ 19
[SENSe:]MIXer:LOSS[:LOW] <Average>
This command defines the average conversion loss to be used for the entire low (first)
range.
Parameters:
<Average> numeric value
Range: 0 to 100
*RST: 24.0 dB
Default unit: dB
Example:
MIX:LOSS 20dB
Manual operation: See "Conversion loss" on page 369
[SENSe:
]MIXer:PORTs <PortType>
This command specifies whether the mixer is a 2-port or 3-port type.
Parameters:
<PortType> 2 | 3
*RST: 2
Example:
MIX:PORT 3
Manual operation: See "Mixer Type" on page 368
[SENSe:]MIXer:RFOVerrange[:STATe] <State>
If enabled, the band limits are extended beyond "RF Start" and "RF Stop" due to the
capabilities of the used harmonics.
Parameters:
<State> ON | OFF
*RST: OFF
Manual operation: See "RF Overrange" on page 368
Conversion Loss Table Settings
The following settings are required to configure and manage conversion loss tables.
[SENSe:]CORRection:CVL:BAND....................................................................................875
[SENSe:]CORRection:CVL:BIAS..................................................................................... 875
[SENSe:]CORRection:CVL:CATAlog?..............................................................................875
[SENSe:]CORRection:CVL:CLEAr................................................................................... 876
[SENSe:]CORRection:CVL:COMMent..............................................................................876
[SENSe:]CORRection:CVL:DATA....................................................................................876
[SENSe:]CORRection:CVL:HARMonic............................................................................. 877
[SENSe:]CORRection:CVL:MIXer....................................................................................877
[SENSe:]CORRection:CVL:PORTs..................................................................................877
[SENSe:]CORRection:CVL:SELect.................................................................................. 878
[SENSe:]CORRection:CVL:SNUMber.............................................................................. 878
Setting Basic Measurement Parameters