EasyManua.ls Logo

Rohde & Schwarz SMW-K144 - Page 180

Rohde & Schwarz SMW-K144
285 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote-Control Commands
R&S
®
SMW-K144
180User Manual 1178.8013.02 ─ 06
Parameters:
<DummyResState> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Filling in the frame with dummy REs"
on page 150.
Manual operation: See "State" on page 41
[:SOURce<hw>]:BB:NR5G:NODE:CELL<ch0>:DUMRes:POWer <Power>
Sets the power of the resource elements filled in with dummy data.
Parameters:
<Power> float
Range: -80 to 10
Increment: 0.001
*RST: 0
Example: See Example "Filling in the frame with dummy REs"
on page 150.
Manual operation: See "Power" on page 42
[:SOURce<hw>]:BB:NR5G:NODE:CELL<ch0>:DUMRes:SCSPacing <SCSPacing>
Selects a combination of the subcarrier spacing (SCS) and the cyclic prefix (CP),
where the available values depend on the "Deployment".
See Table 3-2.
Parameters:
<SCSPacing> N15 | N30 | N60 | N240 | N120
*RST: N30
Example: See Example "Filling in the frame with dummy REs"
on page 150.
Manual operation: See "SC Spacing/CP" on page 42
[:SOURce<hw>]:BB:NR5G:NODE:CELL<ch0>:DUMRes:MODulation <Modulation>
Sets the modulation scheme for the dummy REs.
Parameters:
<Modulation> BPSK | BPSK2 | QPSK | QAM16 | QAM64 | QAM256
*RST: QPSK
Example: See Example "Filling in the frame with dummy REs"
on page 150.
Manual operation: See "Modulation" on page 42
Dummy Resource Elements Commands

Table of Contents

Related product manuals