Operating Instructions Cubis
®
MCA 37
Operating Design
Level 1 Level 2 Level 3 Description
Device
settings
Preload Set preload Sets a preload for technical equipment that
needs to be constantly on the device. In this
case the maximum capacity is reduced
bythe set preload.
Delete preload Deletes the set preload.
Calibration |
adjustment report
Automatic printout
forisoCAL
Activates the data output for the isoCAL
function.
Automatic printout
forcal. | adj. Task
Activates the data output for the calibration
and adjustment function.
Print profile Defines the output device on which
aprintout must be carried out.
minUSP minUSP function If this extension has been activated in the
QAPP Center: Activate or deactivate the
minimum initial weight.
Starting point of
operating range
If this extension has been activated in
theQAPP Center: Define the starting
pointvalue.
Mark weight values
< minUSP as invalid
If this extension has been activated in the
QAPP Center: Activates the identification of
weight values that are less than the minimum
initial weight.
Measurement
uncertainty
Various user entries If this extension has been configured in the
QAPP Center: Activates the dynamic
display of the measurement uncertainty for
the weight value.
Device ID Device ID 1–2 Determine the device identifier.
Electronic signature Sign report If the QAPP extension has been activated:
Activate or deactivate the electronic
signature.
Draft shield Left | right palm key Set the function of the left and right palm
key. Only for devices with a motorized draft
shield.
Automatic mode If the device is equipped with a motorized
draft shield: Closes the draft shield, carries
out the function, and opens the draft
shieldagain.
Sensor mode Only for devices with a motorized draft
shield: Set the operating mode for the
sensor windows.
Left sensor
sensitivityrange
Only for devices with a motorized draft shield:
Define the sensitivity for the left sensor
window of the motion sensor.
Right sensor
sensitivityrange
Only for devices with a motorized draft shield:
Define the sensitivity for the right sensor
window of the motion sensor.