Calibration und Linearization Functions
Select the appropriate menu code to define the access Status for each of the following calibration and linearization functions,
which are activated by holding down the tare control
for a few seconds:
External calibration C.E.
Internal calibration C.I.
Calibration fest C.t.
Internal linearization L.I.
External linearization L.E.
However, if the menu access switch is unlocked ("accessible" Status indicated by -c- after you have accessed the balance
operating menu), the "external calibration" function will be accessible even though you have set the menu code 1 9 2 for
"access denied."
External calibration Code
Accessible * 1 9 1
Access denied 1 9 2
Internal calibration Code
Accessible * 1 10 1
Access denied 1 10 2
Calibration test Code
Accessible * 1 11 1
Access denied 1 11 2
External linearization
The linearization weights to be loaded are displayed on the balance one after the other in increasing order.
External linearization Code
Accessible * 1 12 1
Access denied 1 12 2
Internal linearization Code
Accessible * 1 13 1
Access denied 1 13 2
* = Factory setting