18
Multiple Calibration Mode
The calibration accuracy can be increased. The calibration values calculated from the average of the individual calibration
procedures. You can use the "multiple calibration mode" for both infernal and external calibration. The number of calibration
procedures is indicated in the 3rd place of the application display field (e.g., "C 1 5”).
Important Note:
lf "Err 04" is briefly displayed, the calibration values deviate too much from one another. This means the calibration value
measured is not stored, and the calibration procedure will be repeated.
Multiple calibration mode Code
Off * 1 14 1
On 1 14 2
Fully Automatic Calibration and Linearization
Automatic calibration and linearization Code
Off 1 15 1
Off (with calibration Status)* 1 15 2
Fully automatic calibration on * 1 15 3
Fully automatic calibration and linearization on 1 15 4
Wait Interval until Fully Automatic Calibration Starts
Code
1 minute after stability * 8 9 1
2 minutes after stability 8 9 2
* = Factory setting
* = The symbol "
" flashes in the display until you press the appropriate key to activate one of the calibration functions.