26
GLP/GMP Print or Record
The Research Series Balance can record all completed calibration operations and print out data in compliance with the
requirements of Good Laboratory Practice (GLP).
The balance, interfaced with a Data Printer or a Computer, creates a document that records the date, time, serial number and
model number which makes it possible to clearly trace data to the balance that generated it and the time at which it was
generated.
Application:
Use of the balance in quality assurance Systems in areas subject to compliance with GLP/GMP.
Select the GLP/GMP printout/record mode by setting the appropriate code in the menu of the balance operating program.
GLP/GMP printout/record Code
Off * 8 14 1
Only for calibration functions 8 14 2
Always on 8 14 3
The following are menu code settings which must be selected in order to receive a GLP/GMP printout/record:
Auto. output of reference %/qty and reference weight 7 1 2
or
Auto. output of reference weight only 7 1 3
and
With data ID code 7 2 2
Important Note:
GLP/GMP printouts/records will not be Output if the factory setting 7 1 1 or 7 2 1 "without data l D code" is selected.
In addition, do not select the "Autoprint" data output parameter (menu code 6 1 4 or 6 1 5).
* = Factory setting