Record for Calibration and Linearization Functions
A printout of the recorded data is generated at the end of a calibration or linearization Operation.
The printout of the record can have the following lines:
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
MC1 – Sartorius
: Balance family and manufacturer
Model RC210P
: Balance model
S / N 10205496
: Serial no. of the balance
Id 4-32-1
: ID No. (e.g., for identifying a workstation/operator)
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
-
Date : 30–Sep–91
: Current date
Start : 10:05:30
: Time at which the application started
Cal. : Test
: Calibration mode (in this case, "calibration test")
Diff. : – 0.00013 g
: Data generated during calibration fest
Cal. : Intern
: Calibration mode (in this case, "internal calibration"
Stat. : Complete
: Status comment for calibration or linearization functions
End : 10:05:45
: End of application
Name :
: Field for signature of the operator responsible
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Set. : 200.000000 g
: Calibration weight (only for "external calibration"