EasyManuals Logo

SICK Bulkscan LMS511 User Manual

SICK Bulkscan LMS511
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
OPERATING INSTRUCTIONS| Bulkscan LMS5118014829/ZNC5/2017-06-06| SICK
Subject to change without notice
61
COMMISSIONING
8.5 Scan frequency
The scan frequency is the frequency with which the Bulkscan determines and outputs
measured values at the interface.
Scan frequency Cycle time
35 Hz 28.6 ms
50 Hz 20.0 ms
75 Hz 13.3 ms
Tab. 15: Congurable scan frequencies
IfyouconguretheoutputoftheVolume quota or the Mass quota on a digital
output, then make sure that duration of the pause between pulses is longer than
the cycle time resulting from the scan frequency (see "4.13 Digital outputs“ on page
25).
The selection of the scan frequency affects the energy consumption of the sensor.
System tab, Scan frequency group
8.6 Taking a reference measurement (compensation function)
Depending on the application, systematic errors can occur during the determination of
the volume. In particular the compression of the lower layers of bulk due the mass of
the bulk above will result in reproducible deviations. This systematic error can be calcu-
lated from the volume using a compensation function.
The software calculates this compensation function based on one or more reference
measurements:
Perform up to 10 reference measurements when the load on the conveyor belt varies
in operation.
The best results are achieved if the reference measurements cover the entire mea-
suring range, meaning minimum to maximum belt load.
If you replace a sensor, you can also enter the compensation function directly instead
of the reference measurements.
Requirements for the prevention of measurement errors
Bulk height of the good transported at least 200 mm.
Constant belt speed during the calibration.
Constant bulk properties.
Suitable mounting location and mounting orientation (see "5.2 Selecting the mount-
ing location“ on page 29).
Deformations of the conveyor belt under load < 5 mm.
Reference volume
For the reference measurement you need a reference volume. To determine the refer-
ence volume, a reference measuring system with an accuracy of ±1% with reference
to the end value of the measuring range is required. Examples for suitable reference
measuring systems:
Packing plant for the bulk:
Vessel with straight contours (e.g., container) into which the bulk is transported with
the possibility of subsequently determining the volume.
Note
Note

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK Bulkscan LMS511 and is the answer not in the manual?

SICK Bulkscan LMS511 Specifications

General IconGeneral
BrandSICK
ModelBulkscan LMS511
CategoryAccessories
LanguageEnglish

Related product manuals