Troubleshooting
114 © SICK AG • Subject to change without notice • 8024484/2019-06-06
Display Meaning/possible causes Troubleshooting
eLevel The current attenuation value is below the
warning threshold. The warning threshold
depends on the distance between the
measuringdeviceandthereector.For
recommended attenuation values, see
Page 32, Table 6.
Cleantheopticalinterfaceslikethereec-
tor and the lens.
Reduce the distance between the measur-
ingdeviceandthereector.
Use a distance measuring device with a
larger scanning range. See Page 120,
Chapter 16.
eTemp The internal temperature of the device is
outside the permissible range.
For the permissible ambient temperature,
see Page 119, Chapter 15.8.
Check the ambient temperature. Provide
better ventilation if necessary.
Shield the device from radiated heat, e.g.,
shade the measuring device from direct
sunlight.
At low ambient temperatures, use a heat-
ing system for the device.
At high ambient temperatures, use a
cooling housing.
Table 66: Error messages
13.4 PROFINET IO error messages
Display Meaning/possible causes Troubleshooting
Device error Hardware Check supply voltage and wiring.
For electrical connection Page 36,
Chapter 7.
Switchthesupplyvoltageoandon.
Return the measuring device for repairs.
The internal temperature of the device is
outside the permissible range.
For the permissible ambient temperature,
see Page 119, Chapter 15.8.
Wait for the warm-up phase of the mea-
suring device. Let the measuring device
cool down.
Check the ambient temperature. Provide
better ventilation if necessary.
Shield the device from radiated heat, e.g.,
shade the measuring device from direct
sunlight.
At low ambient temperatures, use a heat-
ing system for the device.
At high ambient temperatures, use a
cooling housing.
Measurement errors Light path between the measuring device
andthereectorinterrupted.
Keep a replacement device ready for use.
The current attenuation value is below the
recommended attenuation value. The recom-
mended attenuation value depends on the
distance between the measuring device and
thereector.For recommended attenua-
tion values, see Page 32, Table 6.
Cleantheopticalinterfaceslikethereector
and the lens.
Lensorreectorcontaminated. Cleantheopticalinterfaceslikethereector
and the lens.
Traversing speed too high. Check maximum traversing speed.