8024484/2019-06-06 • © SICK AG • Subject to change without notice 117
Technical data
Fig. 28: Dimensions of the DL100 Pro distance measuring device
1 Sender optical axis
2 Receiver optical axis
3 Device zero point
4 M5 threaded mounting hole
5 “Status” LED
6 Display
7 Operating elements
15.2 Laser/optics
Light sender Laser diode, red light
Laser protection class 2 in accordance with EN 60825-1 / CDRH
CW modulation ±0.85Posinusoidallymodulated
Maximum output ≤1.9mW
Pulse duration 6.8 ns
Wavelength 655 nm
Frequency ≥90MHz
light spot size Typically 5 mm + (2 mm x distance [m])
Table 68: Laser/optics
15.3 Performance data
Measuring ranges DL100-21xx2213: 0.15 m … 100 m
DL100-22xx2213: 0.15 m … 200 m
DL100-23xx2213: 0.15 m … 300 m
Measurement accuracy Measuringrange0.15mto100m:±2.0mm
Measuringrange0.15m…200m:±2.5mm
Measuringrange0.15m…300m:±3.0mm
Repeatability Measuring range 0.15 m to 100 m: 0.5 mm
Measuring range 0.15 m … 200 m: 1.0 mm
Measuring range 0.15 m … 300 m: 2.0 mm
Initialization time Typically 3 s
Afterreectorloss:<40ms
Resolution Adjustable: 0.1 mm / 0.125 mm / 1.0 mm / 10 mm / 100 mm / free
resolution
Table 69: Performance data