EasyManua.ls Logo

SICK DL100 Pro

SICK DL100 Pro
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8024484/2019-06-06 • © SICK AG • Subject to change without notice 53
Operating the measuring device
8.5.11 “MF1 – Srvice” submenu
InthissubmenuyoucanconguretheMF1multifunctionaloutputasa
serviceoutput.Youcanactivate(On)ordeactivate(O)multipleoptions.
The “Srvice” sub-menu is accessed via the following menu path:
Main menu
Menu Pron more
MFx On MF1 ActSta Functn Srvice
Requirements for display
“more” menu: “Yes” option
“MFx On” menu: “Enable” option
“Functn” parameter: “Srvice” option
Parameters Description
WrnLsr Activates or deactivates the warning message when the aging
of the laser means that the measuring device is nearly due to
be replaced.
Options
On
O
Factory setting
On
WrnLvl Activates or deactivates the warning message when the atten-
uation value is not reached, e.g., due to contamination.
Options
On
O
Factory setting
On
WrnTemp Activates or deactivates the warning message when the
internal temperature of the measuring device is outside the
permissible limits.
Options
On
O
Factory setting
On
WrnPlb Activates or deactivates the warning message when the mea-
sured value is implausible. This may be caused by an errone-
ous measured value, obstruction of the light beam, optical
interference, or electrical interference.
Options
On
O
Factory setting
On

Table of Contents