Troubleshooting
13.2 Warning messages
Display Meaning/possible causes Troubleshooting
NoWrn No warning
wPlb Measured value not plausible.
Light path between the measuring device
andthereectorinterrupted.
Observethelightspotonthereector.The
lightspotmustnotdriftfromthereector.
If necessary, realign the measuring device
andreectororusealargerreector.For
alignment and mounting, see Page 26,
Chapter 6.
Optical interference Remove any optical interference.
Realignthemeasuringdeviceandreec-
tor. For alignment and mounting, see
Page 26, Chapter 6.
wLaser The measurement laser is still ready for op-
eration but is at the end of its service life.
Keep a replacement device ready for use.
wLevel The current attenuation value is below the
recommended attenuation value. The recom-
mended attenuation value depends on the
distance between the measuring device and
thereector.For recommended attenua-
tion values, see Page 32, Table 6.
Cleantheopticalinterfaceslikethereec-
tor and the lens.
Reduce the distance between the measur-
ingdeviceandthereector.
Use a distance measuring device with a
larger scanning range. See Page 120,
Chapter 16
wTemp The internal temperature of the device is ap-
proaching the limit of the permissible range.
For the permissible ambient temperature,
see Page 119, Chapter 15.8.
Check the ambient temperature. Provide
better ventilation if necessary.
Shield the device from radiated heat, e.g.,
shade the measuring device from direct
sunlight.
At low ambient temperatures, use a heat-
ing system for the device.
At high ambient temperatures, use a
cooling housing.
Table 65: Warning messages
13.3 Error messages
Display Meaning/possible causes Troubleshooting
NoErr No fault
ePlb Measured value not plausible.
Light path between the measuring device
andthereectorinterrupted.
Observethelightspotonthereector.The
lightspotmustnotdriftfromthereector.
If necessary, realign the measuring device
andreectororusealargerreector.For
alignment and mounting, see Page 26,
Chapter 6.
Optical interference Remove any optical interference.
Realignthemeasuringdeviceandreec-
tor. For alignment and mounting, see
Page 26, Chapter 6.
eLaser The service life of the measurement laser
has been exceeded.
Replace the measuring device.