EasyManua.ls Logo

SICK FLOWSIC100 Flare-XT

SICK FLOWSIC100 Flare-XT
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
130 FLOWSIC100 Flare-XT · Operating Instructions · 8023761/V 1-0/2020-10 · © SICK Engineering GmbH
Commissioning FLOWSIC100 Flare-XT
Table 16 Selection options
Invert logic: Inverts the logic of the received signal
Raw read: Instantaneous value, without debounce
Debounce: Debounce time (the time a digital input must be constant without status
change)
Alarm on error: In case of an error of the digital input, an error is displayed in the system
status of the FLOWSIC100 Flare-XT
Test mode:
Off: Test mode not active
Permanently on: Test of digital input, permanently on
Permanently off: Test of digital input, permanently off
7.5.6.4 AI/AO
Determine the output values for the analog outputs.
Determine whether an alarm is to be displayed for analog input errors.
Fig. 82 Output via analog output (example)
Lower output value: Minimum output value of the analog output
Upper output value: Maximum output value of the analog output
Test mode: Test mode active
Test value: Test of the output in relation to the selected upper and lower output value
AO check cycle:
A lower and a higher output value can be set for the check cycle. Both values are output at
the start of the check cycle for the time specified in the “Duration of steps” field.
Selection Description
Maintenance Set maintenance condition
Start check cycle Start check cycle of the sender/receiver units
Start AO check cycle Start check cycle of the analog outputs
Start AO check and check
cycle
Start check cycle of analog outputs and sender/receiver
units
Data valid
Overall status for the measuring point; when the control
system signals an incorrect overall status, the Interface Unit
increments the error counters, even when there is no error
on the FLOWSIC100 Flare-XT

Table of Contents

Related product manuals