Installation FLSE100-XT
FLOWSIC100 Flare-XT · Operating Instructions · 8023761/V 1-0/2020-10 · © SICK Engineering GmbH 47
5.6.3
Fitting the spool piece (option)
The spool piece must be mounted in the pipeline so that the arrow marking on it
corresponds to the flow direction.
The flow is output by the measuring system as a positive value when master and slave of
the sender/receiver units for the cross-duct versions are installed according to

p. 18,
§3.3.
Required fitting work
Position the spool piece at the desired section of the pipeline using the hoist.
After attaching the flange bolts, but before tightening, check the correct seating and
alignment of the flange seal.
Align the spool piece so that the offset between inlet pipe, spool piece and outlet pipe is
minimized.
Insert the remaining fixing screws and tighten the nuts crosswise. The applied torque
must not be lower than specified in the project planning.
Install the pressure measurement line between the pressure tapping point (option) and
the pressure sensor (option).
Perform a leak tightness check with suitable means after completion of the installation
work,

p. 74, §5.6.8.4.
WARNING: Danger due to size and mass of the spool piece
Observe the transport information in

p. 44, §5.5.2 !

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK FLOWSIC100 Flare-XT and is the answer not in the manual?

SICK FLOWSIC100 Flare-XT Specifications

General IconGeneral
BrandSICK
ModelFLOWSIC100 Flare-XT
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals