Technical data
FLOWSIC100 Flare-XT · Operating Instructions · 8023761/V 1-0/2020-10 · © SICK Engineering GmbH 155
12 .2
FLSE100-XT sender/receiver units
Table 23 Technical data FLSE100-XT
12.2.1
F1F-S
Power supply
Power supply 12…24 V DC, max. 500 mA
Power input < 1 W
Inputs/outputs
Digital data interfaces 1 x RS485, optically isolated
Approvals
Ex approvals ATEX, IECEx, US/C
Ambient conditions
Temperature range Ignition group IIC T4:
–40 °C ... +70 °C
–50 °C ... +70 °C (optional)
Ignition group IIC T6:
–40 °C ... +55 °C
–50 °C ... +55 °C (optional)
Storage temperature –40 °C ... +70 °C
–50 °C ... +70 °C (optional)
Degree of protection IP 66 / 67
Dimensions
Dimensions (W x H x D) Details, see dimension drawings
Table 24 Technical data F1F-S
Measuring conditions
Operating pressure:
1)
CL150 device flange: 20 bar (g)
PN25 device flange (optional): 20 bar (g)
CL300 device flange (optional). 20 bar (g)
Gas temperature –196 °C ... +280 °C
Ex approvals
IECEx Ex db [ia Ga] IIA T4 Ga/Gb
Ex db [la Ga] IIB T4 Ga/Gb
Ex db [ia Ga] IIC T6/T4 Ga/Gb
Ex ia IIC T6/T4 Ga
ATEX II 1/2G Ex db [ia Ga] IIA T4 Ga/Gb
II 1/2G Ex db [la Ga] IIB T4 Ga/Gb
II 1/2G Ex db [ia Ga] IIC T6/T4 Ga/Gb
II 1G Ex ia IIC T6/T4 Ga
NEC/CEC (US/CA) Class I, Division 1, Group D, T4;
Class I, Zone 1, Ex/AEx d[ia] IIA, T4;
Class I, Division 2, Group D, T4;
Class I, Zone 2, Ex/AEx nA[ia] IIA, T4
Class I, Division 1, Groups C and D, T4;
Class I, Zone 1, Ex/AEx d[ia] IIB, T4;
Class I, Division 2, Groups C and D, T4;
Class I, Zone 2, Ex/AEx nA[ia] IIB, T4
Class I, Division 1, Groups B, C and D, T4;
Class I, Zone 1, Ex/AEx d[ia] IIB + H2, T4;
Class I, Division 2, Groups A, B, C and D, T4;
Class I, Zone 2, Ex/AEx nA[ia] IIC, T4
Installation
Weight ≤ 12 kg (sensor pair)
1)
Temperature-dependent, for details see
p. 165, §12.7
»Derating pressure resistance«