Installation FLSE100-XT
FLOWSIC100 Flare-XT · Operating Instructions · 8023761/V 1-0/2020-10 · © SICK Engineering GmbH 65
5.6.8.1
Calculating the wetted part length wL and geometric installation parameters
Before installation, calculate how deep the sender/receiver units will be pushed into the
pipeline.
The wetted part length depends on:
Nozzle length
Gasket thickness
Ball valve length
Wall thickness
Calculating the wetted part length with the geometry calculator
1 Open the geometry calculator.
2 First select in the “Path number” field of the “Calc length, angle, diameter” tab whether
it is a 1-path or 2-path installation.
3 In the field “Sensor type” (Transducer type), select whether the device is a “Cross duct”
or “Probe” version.
4 Select “retractable” for installations with ball valve and “non retractable” for installa-
tions without ball valve.
5 For cross-duct versions, enter the probe distance “Transducer distance” a determined
for the nozzle installation tool,

p. 54, §5.6.7.2.
6 Enter the required dimensions:
Circumference U
Wall thickness w
Gasket thickness S
Gasket thicknesses D1 and D2
For installation with ball valve: Length of ball valve VL
Angle β: For cross-duct versions, enter the nominal nozzle angle (e.g. 75°, 60°,
45°). For the probe version, measure the installation angle and enter the exact
value (maximum value for measuring the installation angle: ±0,3 °).
The wL (“wetted part length”) and geometric installation parameters are calculated.
NOTICE:
The wall thickness must be determined accurately to 0.1 mm. SICK
recommends using a suitable ultrasonic measuring instrument to determine
the wall thickness.
NOTICE:
The geometric installation parameters in the “Result: Parameters for configu-
ration” area and the wall thickness are required for configuring the measuring
system during commissioning.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK FLOWSIC100 Flare-XT and is the answer not in the manual?

SICK FLOWSIC100 Flare-XT Specifications

General IconGeneral
BrandSICK
ModelFLOWSIC100 Flare-XT
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals