122 FLOWSIC100 Flare-XT · Operating Instructions · 8023761/V 1-0/2020-10 · © SICK Engineering GmbH
Commissioning FLOWSIC100 Flare-XT
7. 1
General information
Before commissioning, the sender/receiver units and the Interface Unit must be
installed and electrically connected.
The display language can be set directly on the device via the display,

p. 122, §7.3.
The field setup wizard in the FLOWgate™ operating software supports commissioning,

p. 123, §7.4.
7. 2 Opening the display protective flap
Loosen the screw on the display protective flap.
Swivel the display protective flap down.
Fig. 73 Display protective flap
7. 3
Setting the display language
The display language can be set directly on the display without login.
Available: English, German, Russian
1 Press ESC to go to the main menu.
2 Using the arrow buttons, navigate to the “Language” menu.
3 Press ENTER to open the “Language” menu.
4 Using the arrow buttons, navigate to the desired language.
5 Press ENTER to confirm the language selection.
The display language is changed.
NOTICE: Display protective flap
Do not remove the display protective flap.
Always keep the display protective flap closed when the display is not in use!
Screw the display protective flap tight after finishing work.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK FLOWSIC100 Flare-XT and is the answer not in the manual?

SICK FLOWSIC100 Flare-XT Specifications

General IconGeneral
BrandSICK
ModelFLOWSIC100 Flare-XT
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals