126 FLOWSIC500 · Operating Instructions · 8025733/1GMJ/V4-2/2022-07 · © SICK Engineering GmbH
Maintenance and meter exchange
Subject to change without notice
5 Check the sealing surfaces on the
adapter (marked green) for
contamination and damage.
6 When the sealing surfaces are
damaged,
e.g. due to corrosion or external
force, the adapter has to be
exchanged.
7 When the adapter is damaged, it has to be removed and a new adapter installed
→ p. 39, §3.3.
Then install the gas flow meter anew, → p. 122, §7.4.9.
8 When the components do not seem to be damaged but leak tightness can not be
established, please contact SICK Customer Service (→ p. 102, §6.1).